U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIFE SPINE, INC. KINETIC SL; SPINAL FIXATION SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIFE SPINE, INC. KINETIC SL; SPINAL FIXATION SYSTEM Back to Search Results
Model Number 1283-055
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/11/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the top screw c4 went thru the locking mechnism of the plate.Illiac creast bone graph, patient seems to be on her way to a solid fusion.
 
Additional Manufacturer Narrative
It was likely the surgeon torqued the screw past the designated diameter that retains the screw within the plate.The distal diameter of the screw hole is tapered in as to retain the screw and indicte to the surgeon that the screw is sufficently inserted.The surgeon may not have identified that the secondary resistance and seemed it was just normal bone resistance, leading to over torquing of the screw that surpassed the retaining ring.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KINETIC SL
Common Device Name
SPINAL FIXATION SYSTEM
Manufacturer (Section D)
LIFE SPINE, INC.
13951 s. quality drive
huntley IL 60142
Manufacturer (Section G)
LIFE SPINE, INC.
13951 s. quality drive
huntley IL 60142
Manufacturer Contact
angela batker
13951 south quality drive
huntley, IL 60142
8478846117
MDR Report Key13529590
Report Number3004499989-2022-00002
Device Sequence Number14469005
Product Code KWQ
UDI-Device Identifier00190837005405
UDI-Public(01)00190837005405
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K073479
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 02/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Other
Device Model Number1283-055
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/13/2022
Initial Report FDA Received Date02/14/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age84 YR
Patient SexFemale
-
-