AMO MANUFACTURING NETHERLANDS TECNIS IOL; EXTENDED DEPTH OF FOCUS INTRAOCULAR LENS Back to Search Results
Model Number ZXT150
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Visual Disturbances (2140); Halo (2227)
Date of Event 12/11/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Ethnicity: unknown as information was asked but it was not provided.The device was not returned for analysis.Therefore, a failure analysis of the complaint device cannot be completed.A review of the device history record, complaint trending, and risk documentation for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that an intraocular lens (iol) was explanted from the patient's right eye due unbearable halos and glare.The patient tried to get used to the halos and glare and even had a yttrium aluminium garnet (yag) laser treatment; however, he just could not drive at night.A vitrectomy, incision enlargement, and sutures were required.Patient outcome is 20/20 sc (without correction), 3 point near vision without glasses (nvsc) equivalent of 20/15.Through follow up, it was learned the patient first observed halos at the 1-day post-operative (op) visit.A non-johnson & johnson lens was successfully implanted as a replacement.No other information was provided.
 
Additional Manufacturer Narrative
Product investigation information was received, and the following fields were updated accordingly: section d-9: device available for evaluation: yes.Section d-9: date returned to manufacturer: 16-mar-2022.Section h-3: device evaluated by manufacturer: yes.Device evaluation: the complaint lens was received inside a specimen cup.Visual inspection under magnification revealed that the lens was covered in dry ovd.The lens was cleaned and visually inspected and no additional issues were observed.No issues related to or could have contributed to visual issues were identified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS IOL
Common Device Name
EXTENDED DEPTH OF FOCUS INTRAOCULAR LENS
Manufacturer (Section D)
AMO MANUFACTURING NETHERLANDS
van swietenlaan 5
groningen, groningen 9728 NX
NL  9728 NX
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key13529785
Report Number3012236936-2022-00395
Device Sequence Number5325728
Product Code POE
UDI-Device Identifier05050474605206
UDI-Public(01)05050474605206(17)230706
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/06/2023
Device Model NumberZXT150
Device Catalogue NumberZXT150U200
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/28/2022
Supplement Date Received by Manufacturer04/01/2022
Initial Report FDA Received Date02/14/2022
Supplement Report FDA Received Date04/07/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/06/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
Patient Weight85 KG
Patient RaceWhite
-
-