• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. MG II BONE SCREW DRILL 3.2MM DIAMETER; INSTRUMENT, KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. MG II BONE SCREW DRILL 3.2MM DIAMETER; INSTRUMENT, KNEE Back to Search Results
Model Number N/A
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/20/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Concomitant devices - nexgen revision block cutter 5mm, 10mm, 15mm, 20mm catalog #: 00598703301; lot #: 60306241.The product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.Multiple mdr reports were filed for this event; please see all reports associated with this event: 0001822565-2022-00397, 0001822565-2022-00398.
 
Event or Problem Description
It was reported that during knee arthroplasty the drill became lodged in the tibial resection guide, bent and fractured.An additional drill and guide were used to complete the procedure.No adverse events were reported as a result of this malfunction.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Visual examination of the returned bone screw drill and block cutter showed signs of repeated use and the drill bit was seized in the block cutter in one of the guide holes.The shaft of the drill bit was identified to be bent and broken.A burr was also identified in another guide hole.Dimensional analysis of the devices were conforming to print specifications where measured.The device history records were reviewed and no discrepancies were identified.A definitive root cause could not be determined with the information provided.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MG II BONE SCREW DRILL 3.2MM DIAMETER
Common Device Name
INSTRUMENT, KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key13533502
Report Number0001822565-2022-00397
Device Sequence Number4227802
Product Code HTW
UDI-Device Identifier00889024186804
UDI-Public(01)00889024186804(10)63651760
Combination Product (Y/N)N
Initial Reporter StateIA
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number00512008500
Device Lot Number63651760
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/26/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/20/2022
Supplement Date Received by Manufacturer04/11/2022
Initial Report FDA Received Date02/15/2022
Supplement Report FDA Received Date04/14/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/09/2017
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Patient SexPrefer Not To Disclose
-
-