| Model Number |
ZRA341S |
| Medical Device Problem Code |
Unintended Movement (3026)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the patient came in with discomfort, and on examination of tooth location (not provided) noticed that one abutment was loose and was removed.Patient returning for replacement.
|
| |
|
Additional Manufacturer Narrative
|
|
Zimmer biomet complaint number (b)(4).Age and date of birth unknown / not provided.Gender unknown / not provided.Patient weight unknown / not provided.Date of event unknown / not provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Abut zirconia 3.5 x 4.5 1 (zra341s) was not returned for investigation.Visual inspection could not be performed.The investigation has been performed based on the available information using the item and lot number along with the applicable device history record (dhr), instructions for use (ifu), and risk file.Functional testing to recreate the reported event could not be performed due to the nature of the device & event.No pre-existing condition was noted, and no per form was provided.Bone density type is unknown.The reported devices were located on unknown tooth sites, and were used for an unknown amount of time.The customer did not provide any pictures or x-rays.Appropriate documentation was reviewed.Dhr review was completed for the subject lot numbers (2020110413).It was confirmed that all operations and inspections were executed as per applicable procedure.No deviations or non-conformances, which could have caused or contributed to the reported event was noted as part of the dhr.Lot was inspected and passed all acceptance criteria by qa.Complaint history review was performed for the reported lot number (2020110413) for similar events and one (1) other relevant complaint was identified (b)(4).Based on the available information, device malfunction could not be verified and the reported event was non-verifiable.The following sections have been updated: b4: date of this report g3: date received by manufacturer g6: checked "follow-up".H2: checked follow-up type.H3: changed "yes" to "no".H6: entered evaluation codes.H10: added manufacturer narrative.
|
| |
|
Event or Problem Description
|
|
No further event information available at the time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
The following sections have been updated: b4: date of this report d4: unique identifier (udi) number g3: date received by manufacturer g6: checked "follow-up" h2: checked follow-up type.H10: added manufacturer narrative.
|
| |
|
Event or Problem Description
|
|
No further event information available at the time of this report.
|
| |
|
Search Alerts/Recalls
|