• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF IQ NATURAL SINGLEPIECE IOL; INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF IQ NATURAL SINGLEPIECE IOL; INTRAOCULAR LENS Back to Search Results
Model Number SN60WF
Medical Device Problem Code Material Opacification (1426)
Health Effect - Clinical Code Visual Disturbances (2140)
Type of Reportable Event Serious Injury
Event or Problem Description
A nurse reported that following an intraocular lens (iol) implant procedure, the patient experienced worsening vision.After examination, iol opacification in the left eye was detected and yag (yttrium aluminum garnet) laser capsulotomy was indicated.However, the lens opacification was not altered even after the procedure was performed, thus maintaining the patient's complaint.
 
Additional Manufacturer Narrative
A sample device was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.There have been no other complaints reported in the lot number.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ NATURAL SINGLEPIECE IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key13534545
Report Number1119421-2022-00282
Device Sequence Number17402907
Product Code HQL
UDI-Device Identifier00380655093177
UDI-Public00380655093177
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
P930014
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2010
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 02/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/30/2013
Device Model NumberSN60WF
Device Catalogue NumberSN60WF.190
Device Lot Number10825180
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/01/2022
Initial Report FDA Received Date02/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/08/2008
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age49 YR
Patient SexMale
-
-