| Model Number |
IPN922810 |
| Medical Device Problem Code |
Gas/Air Leak (2946)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/20/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that "the doctor found cuff leakage when using on patient and changed new one".No patient injury or harm reported.Patient condition reported as "fine".
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that "the doctor found cuff leakage when using on patient and changed new one".No patient injury or harm reported.Patient condition reported as "fine".
|
| |
|
Event or Problem Description
|
|
It was reported that "the doctor found cuff leakage when using on patient and changed new one".No patient injury or harm reported.Patient condition reported as "fine".
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).The actual sample was returned and sent to the manufacturing site for investigation.The manufacturing site reports: "the complaint sample was tested; it could not hold inflation during the test.Another test was performed on the sample and air leak was observed when it was immersed in the water." a device history record review was performed and no relevant findings were identified.Based on the investigation performed, the complaint was confirmed.The root cause was deemed manufacturing related.A capa was opened to further address this issue.
|
| |
|