Model Number 67PFS35
Medical Device Problem Code Inflation Problem (1310)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/09/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during pre-test, the balloon would not properly inflate despite device being completely filled.There was no patient involvement reported.
 
Additional Manufacturer Narrative
One tracheostomy tube was received for evaluation in relation to the reported event.Sample was received f in used condition, without original packaging and with its certificate of safe handling.Functional testing of the device was unable to confirm or replicate the reported event.The pilot balloon was manipulated, and the cuff inflated.After the whole cuff inflated, it was deflated and inflated 4 times, all the times the cuff inflated completely and symmetrically, based on analysis the complaint is not confirmed.Root cause was unable to be determined as the reported event was not confirmed.A dhr review was performed subsequent to the manufacturing of the device and prior to its release.No problems or issues were identified during this dhr review.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX BIVONA FLEXTEND TTS NEO/PED TRACHEOSTOMY TU
Common Device Name
TUBE TRACHEOSTOMY AND TUBE CUFF
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
5700 west 23rd ave
gary IN 46406
Manufacturer (Section G)
SMITHS MEDICAL ASD, INC.
5700 west 23rd ave
gary IN 46406
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key13534820
Report Number3012307300-2022-03383
Device Sequence Number5115194
Product Code JOH
UDI-Device Identifier10351688518613
UDI-Public10351688518613
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K923878
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 02/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Expiration Date04/05/2026
Device Model Number67PFS35
Device Catalogue Number67PFS35
Device Lot Number4136125
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/04/2021
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/12/2021
Initial Report FDA Received Date02/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/24/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-