| Model Number |
FS116 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/30/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned to smiths medical.A sample was test for leakage by manually introducing water using a syringe.The sample present leakage by one of the side of the filter.It was observed that the filter is open in one side where should be welded.It present damaged which may indicate a wrong handling of the part.Dhr review was executed and no discrepancies were observed during the manufacturing of reports lot 4068818.The lot used of filter p1123-07 was 6447/20, no defects were reported on this lot of filter.Complaint is confirmed due to sample return present leakage at filter, however root cause is not attributable to manufacturing.Supplier notification was issued on 29-jun-21 to supplier borla.
|
| |
|
Event or Problem Description
|
|
It was reported that the picu has had "7-8+" occurrences of cracked/leaking filters from the same lot, so 9 files created.No patient injury was reported.
|
| |
|
Search Alerts/Recalls
|