ALLERGAN (COSTA RICA) STYLE 110 SILICONE GEL FILLED BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
|
Back to Search Results |
|
| Model Number |
110-240 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Healthcare professional reported right side exchange from textured to smooth implants due to the patient's concern with the product, rupture and any creases associated with hole.Device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture.Device evaluation: analysis of the returned device identified: underweight, crease fold and opening on posterior.A visual and microscopic analysis was performed which identified: crease sharp opening on posterior and wear abrasion.Base on the device analysis the final assessment is: crease sharp opening on posterior assessed as fold flaw opening wrinkles (creases) are observed but have no relationship with manufacturing.
|
| |
|