Model Number BA22-90/I16-30
Medical Device Problem Codes Off-Label Use (1494); Separation Problem (4043)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 02/07/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device remains implanted in the patient; therefore, it will not be returned for evaluation.Patient medical records and imaging studies will be requested for further evaluation by the clinical specialist.If additional information pertinent to the incident is obtained, a follow-up report will be submitted.Device iteration is afx with duraply.
 
Event or Problem Description
The patient was initially implanted with a bifurcated stent graft and a suprarenal stent graft extension to treat an abdominal aortic aneurysm (aaa).Approximately 6 years post initial procedure, a type 3a endoleak was identified during a routine follow up.An intervention has been scheduled.The patient was reported as doing fine.
 
Event or Problem Description
The patient was initially implanted with a bifurcated stent graft and a suprarenal stent graft extension to treat an abdominal aortic aneurysm (aaa).Approximately 6 years post initial procedure, a type 3a endoleak was identified during a routine follow up.An intervention has been scheduled.The patient was reported as doing fine.After the initial report clinical assessment determined that there was evidence to reasonably suggest a type ii endoleak (non-device related) of the lumbar artery occurred that was not included in the event as reported.The type ii endoleak was discovered during a review of the 69-month post index computerized tomography scan.
 
Additional Manufacturer Narrative
The reported adverse event/incident was investigated in alignment with endologix operating procedures and work instructions.Where possible, it is endologix practice to make at least three good faith efforts to retrieve a reported adverse event/incident related device as well as medical records and medical imaging.An evaluation of the manufacturing record was completed.A review of the part number/lot number combination needed for device identification shows the device to have been properly manufactured and released in accordance with the device master record.An evaluation of the device could not be completed.The device was not returned to endologix for evaluation because it remains implanted.A clinical evaluation of the adverse event/incident was completed.An examination of medical records and/or medical imaging received by endologix shows the type iiia endoleak of the aortic components and sac growth of 17.7mm are confirmed.This is consistent with the reported adverse event/incident.The clinical evaluation also shows reasonable evidence to suggest a type ii endoleak of the lumbar artery occurred that was not included in the event as reported.These findings were discovered during an examination of the 69-month post index computerized tomography scan.The most likely causation for type iiia endoleak of the aortic components and sac growth of 17.7mm is anatomy.The superior stent margin of the bifurcated stent graft was positioned in the aortic angulation, which most likely contributed to the type iiia endoleak.Procedure related harms for this complaint could not be determined.During the investigation, endologix found reasonable evidence to suggest the device was used off-label.The iliac diameters were off-label at the index procedure; however, this did not appear to contribute to the reported events.The final patient status was reported as being fine.No additional investigation of this reported adverse event/incident is planned.However, should additional information relevant to the investigation outcome become available, a follow-up report will be submitted.Endologix will continue to monitor this and similar adverse events/incidents.Device iteration is afx with duraply.B5: describe event or problem - updated.G3: awareness date ¿ updated.H6: investigation finding codes - remove code 3233.H6: investigation conclusion codes - remove code 11.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AFX
Common Device Name
BIFURCATED STENT GRAFT
Manufacturer (Section D)
ENDOLOGIX
2 musick
irvine CA 92618
Manufacturer (Section G)
ENDOLOGIX
2 musick
,
irvine CA 92618
Manufacturer Contact
gary kirchgater
2 musick
,
irvine, CA 92618
8009832284
MDR Report Key13535854
Report Number2031527-2022-00034
Device Sequence Number19890620
Product Code MIH
UDI-Device Identifier00818009010223
UDI-Public(01)00818009010223(17)171120
Combination Product (Y/N)N
Initial Reporter StateAL
Initial Reporter CountryUS
PMA/510(K) Number
P040002
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,Company Representative
Initial Reporter Occupation Physician
Remedial Action Recall,Other
Type of Report Initial,Followup
Report Date (Section B) 02/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/20/2017
Device Model NumberBA22-90/I16-30
Device Lot Number1265951-002
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/07/2022
Supplement Date Received by Manufacturer03/21/2022
Initial Report FDA Received Date02/15/2022
Supplement Report FDA Received Date04/11/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/16/2014
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-0006-2019
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
AFX, VELA SUPRARENAL, LOT#1252431028.
Outcome Attributed to Adverse Event Other;
Patient Age84 YR
Patient SexFemale
-
-