SMITH & NEPHEW, INC. MOTOR DRIVE UNIT HAND CNTRL PWRMX EL; ARTHROSCOPE Back to Search Results
Model Number 72200616
Device Problems Overheating of Device (1437); Mechanical Jam (2983)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/26/2022
Event Type  malfunction  
Event Description
It was reported that the motor drive unit had no movement and it was heating up.No case reported; therefore, there was no patient involvement.
 
Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Manufacturer Narrative
Internal complaint reference case (b)(4).H10 h3, h6: the reported device was received for evaluation.It was determined the device contributed to the reported event.A visual inspection found a broken lanyard.A functional evaluation revealed a handpiece stall error.No overheating was noted.Further evaluation revealed a defective power cord.It was determined that the power cord has shorted internal wiring.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failures were documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.The complaint was confirmed, and the root cause was associated with a short circuit in the power cord.Factors which can contribute to a shorted power cord include rolling a heavy cart over the cord or excessive bending of the cord.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MOTOR DRIVE UNIT HAND CNTRL PWRMX EL
Type of Device
ARTHROSCOPE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
Manufacturer (Section G)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key13536167
MDR Text Key287923959
Report Number1643264-2022-00052
Device Sequence Number1
Product Code HRX
UDI-Device Identifier03596010555472
UDI-Public03596010555472
Combination Product (y/n)N
Reporter Country CodeTC
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 04/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number72200616
Device Catalogue Number72200616
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/27/2022
Initial Date FDA Received02/15/2022
Supplement Dates Manufacturer Received04/05/2022
Supplement Dates FDA Received04/06/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-