DATASCOPE CORP. - MAHWAH CS300 INTRA-AORTIC BALLOON PUMP, FRENCH, 110V; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0998-00-0446-61
Medical Device Problem Code Device Contamination with Body Fluid (2317)
Health Effect - Clinical Codes Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/20/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Type of investigation not yet determined : the customer has been presented with a quote for the cost estimate for repairs.Additional information is being requested with regards to the patient complication.Additionally, a report of user error was provided and we are seeking additional clarification.A supplemental report will be submitted upon completion of our investigation or if additional information is provided.
 
Event or Problem Description
It was reported by the customer that the cs300 intra-aortic balloon pump (iabp) unit had blood backed all the way through the tubing and potentially into the machine.It was reported that the patient had a complication that was not due to the cs300 device.
 
Additional Manufacturer Narrative
Corrected fields: h6 (type of investigation, component codes).Testing of actual/suspected device (10/3233): a getinge field service engineer (fse) was dispatched to investigate.The fse observed traced of blood in the purge line up to the tank tubing after the purge valve assembly.The fse replaced all blood contaminated parts which include the pneumatic component male luer, reservoir assembly, fill manifold bracket, purge valve assembly, fill manifold assembly, block single transducer, blood detect tubing, condensation removal module, and safety disk.During testing, the fse observed that the drive manifold was leaking.This unrelated issue will be documented on a separate medwatch report.The iabp unit remains out of use pending additional repairs.A supplemental report will be submitted when additional information is provided.
 
Event or Problem Description
N/a.
 
Event or Problem Description
It was reported that during use on a patient, the cs300 intra-aortic balloon pump (iabp) had blood backed all the way through the tubing and potentially into the machine.There was no patient harm and no adverse event was reported.
 
Additional Manufacturer Narrative
Additional information: (event site postal code: (b)(4)).
 
Event or Problem Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CS300 INTRA-AORTIC BALLOON PUMP, FRENCH, 110V
Common Device Name
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key13536784
Report Number2249723-2022-00302
Device Sequence Number5075529
Product Code DSP
UDI-Device Identifier10607567108292
UDI-Public10607567108292
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K063525
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date (Section B) 03/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number0998-00-0446-61
Device Catalogue Number0998-00-0446-61
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/20/2022
Supplement Date Received by Manufacturer04/04/2022
04/19/2022
06/05/2022
02/28/2024
Initial Report FDA Received Date02/15/2022
Supplement Report FDA Received Date04/29/2022
05/12/2022
12/14/2022
03/12/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/24/1998
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
INTRA-AORTIC BALLOON CATHETER
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-