| Model Number |
3501490 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Deformity/ Disfigurement (2360)
|
| Date of Event |
12/01/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: deflation.Manufacturer¿s reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a (b)(6)-year-old female patient underwent a breast reconstruction surgery with implantation of a 575cc mentor smooth round spectrum saline breast implant prosthesis.Post-operatively, the patient observed her left breast implant prosthesis to be deflated.As a result, the patient underwent removal and replacement with a 275cc mentor smooth round moderate plus profile saline breast implant prosthesis on (b)(6) 2022.
|
| |
|
Additional Manufacturer Narrative
|
|
On (b)(6) 2022, mentor became aware that an error was submitted in the initial report.The initial reporter should have been indicated as not a health professional.Manufacturer¿s reference number: (b)(4) in the initial report, "e2.Health professional?', should have been no.
|
| |
|
Additional Manufacturer Narrative
|
|
On april 11, 2022, mentor completed an evaluation on the returned device.Mentor conducted a visual inspection and leak testing of the device.Visual analysis of the returned sample revealed that the breast implant had a crease/fold on the posterior view.Leak testing was performed, according to the mentor procedure, and it identified a tear within the crease/fold measuring approximately 0.1 cm.The evaluation determined that the cause of the rupture is consistent with normal wear.Deflation can occur at any time after implantation, but it is more likely to occur the longer the implant is implanted.Creating wrinkles or folds should be avoided during implantation or other procedures as it can result in deflation at a later time.Breast implants may also simply wear out over time.Manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
On march 18, 2022, the mentor failure analysis lab has received the device for evaluation.The analysis has begun but is not complete at this time.When the investigational analysis has been completed, a supplemental report will be submitted.Manufacturer¿s reference number: (b)(4).
|
| |
|