• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON INC. PDSII CLR 18IN 3-0 S/A PS-2 PRM MP; SUTURE, SURGICAL, ABSORBABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. PDSII CLR 18IN 3-0 S/A PS-2 PRM MP; SUTURE, SURGICAL, ABSORBABLE Back to Search Results
Model Number Z497G
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a patient underwent an unknown procedure on an unknown date and suture was used.It reported that the device is breaking too easy.No adverse patient consequences were reported.Additional information was requested.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent.What was the initial procedure? what is the procedure & event date? could you please confirm if the issue occurred before used it in patient or during use? please clarify please confirm what was breaking, in event description it is unclear "the device is breaking"? a manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.Event related to mw # 2210968-2022-01076 and mw # 2210968-2022-01077.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Date sent to the fda: 3/11/2022.H6 component code: g07002 no device problem found.H3 investigational narrative: one opened box with twelve packets of product code z497 were returned to ethicon inc for analysis with the packaging closed.Upon initial inspection to the sample no external damages were observed on the packet.In order to evaluate the conditions of the returned samples, the packets were opened, and no defects were detected.The swage and attachment area were noted to be as expected.The needles were intact and no damages, breakage on body, tip or swage area were observed during evaluation.Functional test was performed, to needles by resistance and met the requirements.As part of our quality process, the manufacturing records of this lot-serial number were reviewed, and the manufacturing standards were met prior to the release of this batch.As part of ethicon quality process, all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PDSII CLR 18IN 3-0 S/A PS-2 PRM MP
Common Device Name
SUTURE, SURGICAL, ABSORBABLE
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.-JUAREZ
avenida de las torres 7125
col salvacar
ciudad juarez 32604
MX   32604
Manufacturer Contact
elba bello
1000 route 202
raritan, NJ 08869
MDR Report Key13538314
Report Number2210968-2022-01075
Device Sequence Number4063500
Product Code NEW
UDI-Device Identifier10705031061231
UDI-Public10705031061231
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
N18331
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2026
Device Model NumberZ497G
Device Catalogue NumberZ497G
Device Lot NumberRKMACH
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/14/2022
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/20/2022
Supplement Date Received by Manufacturer02/16/2022
Initial Report FDA Received Date02/15/2022
Supplement Report FDA Received Date03/11/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/01/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-