| Model Number |
BEA28-120/I20-40 |
| Medical Device Problem Codes |
Off-Label Use (1494); Material Puncture/Hole (1504); Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
02/01/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The patient was initially treated for an abdominal aortic aneurysm (aaa), with the implant of an afx2 bifurcated stent graft, afx vela infrarenal stent graft, and an afx suprarenal stent graft.Approximately four (4) years post initial procedure, the patient underwent a computed tomography (ct) scan where an endoleak (indeterminate origin) was identified.The patient is currently in stable condition pending a re-intervention.
|
| |
|
Additional Manufacturer Narrative
|
|
The devices involved in this event will not be returned for evaluation and remain implanted in the patient.Patient medical records and imaging studies will be requested for further evaluation by a clinical specialist.If additional information pertinent to the incident is obtained, a follow-up report will be submitted.Device iteration is afx2.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported adverse event/incident was investigated in alignment with endologix operating procedures and work instructions.Where possible, it is endologix practice to make at least three good faith efforts to retrieve a reported adverse event/incident related device as well as medical records and medical imaging.An evaluation of the manufacturing record was completed.A review of the part number/lot number combination needed for device identification shows the device to have been properly manufactured and released in accordance with the device master record.An evaluation of the device could not be completed.The device was not returned to endologix for evaluation because it remains implanted.A clinical evaluation of the adverse event/incident was completed.An examination of medical records and/or medical imaging received by endologix shows the indeterminate endoleak was found to be a type ia endoleak and a type iiib endoleak of the distal main body.This is moderately consistent with the reported adverse event/incident.The distal type iiib endoleak was seen on a 2018 computerized tomography scan.The most likely causation for the type ia endoleak is user related.Procedure related harms for this complaint could not be determined.During the investigation, endologix found reasonable evidence to suggest the device was used off-label.The use of a non-endologix stent in the proximal neck (off label - concomitant use with products outside the ifu) this likely contributed to the type ia endoleak.The final patient status was reported to be in stable condition.No additional investigation of this reported adverse event/incident is planned.However, should additional information relevant to the investigation outcome become available, a follow-up report will be submitted.Endologix will continue to monitor this and similar adverse events/incidents.Device iteration is afx2.
|
| |
|
Event or Problem Description
|
|
The patient was initially treated for an abdominal aortic aneurysm (aaa), with the implant of an afx2 bifurcated stent graft, afx vela infrarenal stent graft, and an afx suprarenal stent graft.Approximately four (4) years post initial procedure, the patient underwent a computed tomography (ct) scan where an endoleak (indeterminate origin) was identified.The patient is currently in stable condition pending a re-intervention.After the previously filed report, additional information was received reporting that reintervention was completed with implant of an afx2 bifurcated stent graft and an ovation ix extender.The endoleak was successfully sealed.
|
| |
|