Model Number SRLP-340
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code Failure of Implant (1924)
Type of Reportable Event Serious Injury
Event or Problem Description
Healthcare professional reported left side rupture.Device remains implanted.
 
Additional Manufacturer Narrative
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture.
 
Event or Problem Description
Device has been explanted.
 
Additional Manufacturer Narrative
Device evaluation: analysis of the returned device identified: weight to spec, crease fold, wear abrasion, observed split in patch area, it is out of specification per qa 199.04 and white particles in the shell.Air voids in the gel observed after autoclave cycle.The device is no rupture.Base on the device analysis the final assessment is: split in patch area, assessed as a workmanship.
 
Event or Problem Description
Healthcare professional reported left side rupture.The device has been replaced.
 
Additional Manufacturer Narrative
Further investigation (fi): the review of the documentation associated to the manufacturing process indicates that all devices with work order #(b)(4) were released in accordance with abbvie¿s procedures, and no anomalies were found in the documents or in the personnel training related to the reported event.According to the device analysis performed in the laboratory, it was observed split in patch area, it is out of specification per qa 199.04 which is not related to the reported complaint.A review of the current risk documents was performed and the event of split in patch is a known event, for which manufacturing process controls and inspections are stablished to reduce the incidence of this defect.During the trend review of all split in patch complaints for gel breast implants for the period of may 2020 through apr 2022, was noted an outlier in jun 2020.The additional analysis performed shows all that results met the acceptance criteria and no issues, deviations or non-conformances were found which could be associated to the split in patch event, so, no additional actions are deemed required at this time.The issue with split in patch will continue to be monitored and corrective action taken in the future if deemed appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INSPIRA SMOOTH SILICONE GEL FILLED BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
Manufacturer (Section G)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
Manufacturer Contact
terry ingram
12331-a riata trace parkway
building 3
austin, TX 78727
8479366324
MDR Report Key13538664
Report Number9617229-2022-01653
Device Sequence Number5147850
Product Code FTR
UDI-Device Identifier10888628006928
UDI-Public10888628006928
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P020056
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/17/2020
Device Model NumberSRLP-340
Device Catalogue NumberSRLP-340
Device Lot Number2776906
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/25/2022
Supplement Date Received by Manufacturer05/11/2022
06/24/2022
Initial Report FDA Received Date02/15/2022
Supplement Report FDA Received Date06/08/2022
07/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/17/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
Patient Weight58 KG
-
-