BOSTON SCIENTIFIC CORPORATION REBEL; STENT, CORONARY Back to Search Results
Model Number 10275
Medical Device Problem Codes Break (1069); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/04/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that shaft break occurred.A 28 x 4.00 rebel bms stent was advanced for treatment.However, while advancing it over the wire, the shaft break.The device was pulled out and the procedure was completed with another of the same device.There were no patient complications reported.
 
Event or Problem Description
It was reported that shaft break occurred.A 28 x 4.00 rebel bms stent was advanced for treatment.However, while advancing it over the wire, the shaft break.The device was pulled out and the procedure was completed with another of the same device.There were no patient complications reported.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: returned product consisted of a rebel bms balloon catheter.The device was visually and microscopically examined.At 83.4cm from the strain relief there was a hypotube separation.The separated ends were ovaled in shape indicating the device was kinked prior to separation.There were numerous hypotube kinks, and contrast was present in the inflation lumen and the balloon.The balloon was tightly folded, and the stent was found to be intact and in place.Product analysis confirmed the reported break as there was separation of the device.There were also numerous unreported kinks.
 
Event or Problem Description
It was reported that shaft break occurred.A 28 x 4.00 rebel bms stent was advanced for treatment.However, while advancing it over the wire, the shaft break.The device was pulled out and the procedure was completed with another of the same device.There were no patient complications reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REBEL
Common Device Name
STENT, CORONARY
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key13538708
Report Number2134265-2022-01553
Device Sequence Number17474000
Product Code MAF
UDI-Device Identifier08714729844013
UDI-Public08714729844013
Combination Product (Y/N)N
Initial Reporter StateIA
Initial Reporter CountryUS
PMA/510(K) Number
P130030
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 03/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/16/2023
Device Model Number10275
Device Catalogue Number10275
Device Lot Number0026366291
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/04/2022
Supplement Date Received by Manufacturer03/09/2022
Initial Report FDA Received Date02/15/2022
Supplement Report FDA Received Date03/17/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/16/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1