• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SPECTRANETICS ANGIOSCULPT OTW PTA; CATHETER, PTA, CUTTING/SCORING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SPECTRANETICS ANGIOSCULPT OTW PTA; CATHETER, PTA, CUTTING/SCORING Back to Search Results
Model Number 2092-6040
Medical Device Problem Codes Material Rupture (1546); Improper or Incorrect Procedure or Method (2017)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/29/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The patient's dob or age at time of event, gender, weight, ethnicity, and race are unknown.This information was not available from the facility.Patient information regarding relevant tests/laboratory data or medical history are unknown.This information was not available from the facility.Foreign: (b)(6).The angiosculpt catheter was returned for evaluation.Visual inspection found no unusual characteristics of the device.During functional testing, using an indeflator filled with green color dye water, the device was pressurized up to 10 atm and a leak on the balloon was present.Under microscopic inspection, a longitudinal tear on the proximal portion of the balloon was observed.The user failed to follow instructions.An undersized introducer sheath (5f) was used.The instructions for use indicates the device is compatible with a 6f introducer sheath as indicated on the product label.In addition, the ifu warns at least 99.9% of the balloons (with a 95% confidence level) will not burst at or below the rbp.
 
Event or Problem Description
The angiosculpt device was used to treat a slightly calcified peripheral lesion.During inflation, the balloon could not be inflated over 10 atm due to a leak on the hub.The procedure was completed with a new angiosculpt device.No patient injury reported.During the returned device evaluation, a balloon rupture was observed.This product problem is being submitted conservatively because the balloon ruptured below rbp.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANGIOSCULPT OTW PTA
Common Device Name
CATHETER, PTA, CUTTING/SCORING
Manufacturer (Section D)
SPECTRANETICS
5055 brandin court
fremont CA 94538
Manufacturer (Section G)
SPECTRANETICS
5055 brandin court
fremont CA 94538
Manufacturer Contact
ana tan
5055 brandin court
fremont, CA 94538
MDR Report Key13539095
Report Number3005462046-2022-00010
Device Sequence Number3734427
Product Code PNO
UDI-Device Identifier00813132021436
UDI-Public00813132021436
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K100303
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 01/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/08/2024
Device Model Number2092-6040
Device Catalogue Number2092-6040
Device Lot NumberG21070007
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/19/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/26/2022
Initial Report FDA Received Date02/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/08/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
BOSTON SCIENTIFIC: 0.018" FUGA GUIDEWIRE; MEDIKIT: 5F ULTRA HIGH FLOW INTRODUCER SHEATH
Patient SexUnknown
-
-