Reported event: an event regarding loosening involving a dall miles cable was reported.The event was confirmed via clinician review.Method & results: product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device was not returned.Clinician review: a review of the provided medical information by a clinical consultant indicated: this inquiry reports a revision of a primary hip replacement immediately after surgery when penetration of the lateral cortex by the femoral stem was noted on the immediate post-operative x-ray.It also reports another surgery due to persistent pain following the first revision where a loose cable was removed.It also reports a second revision due to loosening of the femoral stem.I can confirm that each of these events took place since i was able to review operation notes and x-rays.I was not provided with the x-ray that showed the penetration of the lateral femur but it was documented by the operating surgeon in his operation report.As for the possible root causes of this events, regarding the issue of penetration of the lateral cortex, this is a technical surgical issue that was recognized in a timely fashion and dealt with appropriately.Regarding the operation to remove a loose cable, this can often occur despite the cable being tight and secure after insertion.Causes are multifactorial and can be due to local changes about the bone, inflammation or infection and surgical technique.Regarding the revision for loosening of the femoral stem, causes are multifactorial, including surgical technique, inflammation or infection, and patient factors.Product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusions: it was reported that the patient's hip construct was revised due to pain.Clinician review indicated that regarding the operation to remove a loose cable, this can often occur despite the cable being tight and secure after insertion.Causes are multifactorial and can be due to local changes about the bone, inflammation or infection and surgical technique.The exact cause of the event could not be determined.No further investigation for this event is possible at this time.If devices and / or additional information become available to indicate further evaluation is warranted, this record will be reopened.
|
|
Patient stated she had a left tha on (b)(6) 2016.Two hours after her surgery the patient was brought back to the operating room and was told the piece failed and was revised.Patient stated she was experiencing pain and have seen her surgeon but was told that everything was fine.The patient sought a second opinion on (b)(6) 2018 and was revised due to her x-ray findings.Patient continued to experience pain and was revised on (b)(6)2019.Patient is still experiencing excruciating pain.Patient has been in and out of the hospital for the past 5 years and is seeking compensation for all her suffering.This pi is for revision1.Update on 21-jan-2022 based on med review: an operation note on (b)(6) 2018.The postoperative diagnosis was "painful hardware left hip, removal of hardware left hip, we are removing a cable off the femur".No signs of infection were noted a normal amount of joint fluid was seen with no signs of purulence.Frozen section came back not consistent with infection.The cable was loose and removed.A repair of the abductor mechanism was carried out.
|