C.R. BARD, INC. (BASD) -3006260740 PONSKY PEG KIT, 20F; FEEDING TUBE Back to Search Results
Model Number 006035
Medical Device Problem Codes Fracture (1260); Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/29/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that sometime post feeding device placement procedure, the internal bumper of the device allegedly detached and the tube came off when the device was tried to be used, the internal bumper was reported to have been excreted with the stool.The procedure was completed using another device.There was no reported patient injury.
 
Additional Manufacturer Narrative
As the lot number for the device was not provided, a review of the device history records could not be performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.
 
Additional Manufacturer Narrative
H10: manufacturing review: a manufacturing review was not requested as the lot number reported is unknown.Investigation summary: one ponsky dual port feeding dome was returned for evaluation.Gross visual and microscopic visual evaluations were performed on the returned device.During visual evaluation, a complete circumferential break was noted on the dome, however the remaining feeding tube or adaptor was not returned for evaluation.Therefore, the investigation is inconclusive for the reported fracture and material separation issues as the returned separated dome itself is not sufficient enough to confirm/unconfirm the reported events and as well the exact circumstances at the time of the reported event are unknown.The definitive root cause could not be determined based upon available information.Labeling review: a review of product labeling documentation (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that sometime post feeding device placement procedure, the internal bumper of the device allegedly detached and the tube came off when the device was tried to be used, the internal bumper was reported to have been excreted with the stool.It was further reported that the device has been removed and the procedure was completed using another device.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PONSKY PEG KIT, 20F
Common Device Name
FEEDING TUBE
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD SHANNON LIMITED -3005636544
san geronimo industrial park
lot #1, road #3, km 79.7
humacao PR 00791
Manufacturer Contact
judy ludwig
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key13539955
Report Number3006260740-2022-00328
Device Sequence Number17257462
Product Code KNT
UDI-Device Identifier00801741081187
UDI-Public(01)00801741081187
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K915841
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number006035
Device Catalogue Number006035
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/02/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 01/20/2022
Supplement Date Received by Manufacturer02/28/2022
Initial Report FDA Received Date02/16/2022
Supplement Report FDA Received Date03/03/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexFemale
-
-