Model Number CD1211-36Q
Medical Device Problem Code Charging Problem (2892)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/07/2022
Type of Reportable Event Serious Injury
Event or Problem Description
Related manufacturer report number: 2017865-2022-03436.During an in clinic follow up, an ocd (over current detection) alert was received on the device indicating a possible high voltage lead issue on the right ventricular (rv) lead.Technical support was contacted and noted shocks had not been delivered due to high defibrillation impedance on the rv lead.A lead fracture was suspected.Technical support recommended lead and device replacement.The rv lead and the device were explanted and replaced to resolve the event.The patient was stable.
 
Additional Manufacturer Narrative
The reported field event of over current detection (ocd) could not be reproduced in the lab.Visual inspection did not reveal any anomaly.The device was above elective replacement indicator (eri) upon receipt.Telemetry, impedance, sensing, pacing and high voltage (hv) output functions of the device were tested and found to be normal.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CURRENT PLUS VR, DF-4 CONNECTOR
Common Device Name
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key13541201
Report Number2017865-2022-03437
Device Sequence Number14407589
Product Code NVZ
UDI-Device Identifier05414734503716
UDI-Public05414734503716
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
P910023
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/30/2015
Device Model NumberCD1211-36Q
Device Catalogue NumberCD1211-36Q
Device Lot Number4604544
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/07/2022
Supplement Date Received by Manufacturer04/08/2022
Initial Report FDA Received Date02/16/2022
Supplement Report FDA Received Date04/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/08/2014
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DURATA STS OPTIM ACTIVE FIXATION, DF-4 CONNECTOR
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-