• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS COBAS C111; CLINICAL CHEMISTRY ANALYZER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS COBAS C111; CLINICAL CHEMISTRY ANALYZER Back to Search Results
Model Number C111
Medical Device Problem Code Low Test Results (2458)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/25/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The customer stated they had to run qc nine times at startup because they were repeatedly out of range for "one level or the other".The customer then calibrated twice before running the samples.The field service engineer (fse) ran a precision check, ise performance check, and calibration and qc check.All checks met acceptance criteria.The investigation is ongoing.
 
Event or Problem Description
There was a complaint of questionable ise sodium results for 2 patients tested with a cobas c111 module.The questionable results were not reported outside the laboratory.Patient 1¿s initial result was 118 mmol/l; the repeat was 135 mmol/l.Patient 2¿s initial result was 122 mmol/l; the repeat was 137 mmol/l.The customer believes the repeated result to be correct.The sodium electrode used was lot 21512647 and had an installation date of (b)(6) 2021.
 
Additional Manufacturer Narrative
The customer's calibration and qc results were ok.There was no indication of a performance issue of reagent or instrument.The system's alarm trace was requested but not provided.The investigation did not identify a product problem.The cause of the event could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COBAS C111
Common Device Name
CLINICAL CHEMISTRY ANALYZER
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE INSTRUMENT CENTER AG TEGIMENTA
forrenstrasse
na
rotkreuz 6343
SZ   6343
Manufacturer Contact
michael leslie
9115 hague road
na
indianapolis, IN 46250
3175214343
MDR Report Key13541340
Report Number1823260-2022-00393
Device Sequence Number4456809
Product Code JJE
UDI-Device Identifier04015630924769
UDI-Public04015630924769
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K071211
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberC111
Device Catalogue Number04777433001
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/27/2022
Supplement Date Received by Manufacturer03/07/2022
Initial Report FDA Received Date02/16/2022
Supplement Report FDA Received Date03/19/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age22 YR
Patient SexFemale
-
-