| Model Number |
391020 |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/01/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Combination product: yes.
|
| |
|
Event or Problem Description
|
|
An orsiro drug-eluting stent system was selected for treatment of a severely calcified lesion (90 percent stenosis degree) in the mildly tortuous rca.After rotablation, several attempts were made to place the orsiro stent through the 7f guidezilla, but the stent could not pass through the lesion.When the stent was removed for additional dilatation, the mid of the stent was rolled up.Another stent was successfully implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Combination product: yes.
|
| |
|
Event or Problem Description
|
|
An orsiro drug-eluting stent system was selected for treatment of a severely calcified lesion (90 percent stenosis degree) in the mildly tortuous rca.After rotablation, several attempts were made to place the orsiro stent through the 7f guidezilla, but the stent could not pass through the lesion.When the stent was removed for additional dilatation, the mid of the stent was rolled up.Another stent was successfully implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Combination product: yes.The returned product was subjected to a technical analysis and the corresponding product release documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for this event.The technical investigation of the returned instrument revealed that the stent is severely deformed in its center.Several stent struts are lifted and everted.Microscopic analysis revealed stent imprints on the exposed balloon surface, indicating that the lifted struts were initially crimped on the balloon.Review of the product release documentation confirmed that the instrument was manufactured according to specifications and passed all in-process and final inspections.As a part of final inspection every stent system undergoes visual inspection to ensure correct embedding and homogeneous crimping of the stent.Based on the conducted investigations, no manufacturing or material related root cause could be determined.
|
| |
|
Search Alerts/Recalls
|