Model Number 100/897/080
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Codes Respiratory Insufficiency (4462); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/31/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device was unable to be evaluated as it was not returned.If the device is received in the future an evaluation will be performed.The reported event was unable to be confirmed or replicated.A cause of the reported event could not be determined at this time.A dhr review was unable to be completed as no lot information was provided.
 
Event or Problem Description
It was reported that th tracheostomy tube experienced a disconnection between the reinforced section and 30mm connector.This is an undetachable part of the tube, it appears to have become loose due to humidification.The tracheostomy tube had been in for 19 days.This resulted derecruitment of both lungs and desaturations, which was resolved by the placement of a new tracheostomy tube.
 
Additional Manufacturer Narrative
The device was unable to be evaluated as it was not returned.If the device is received in the future an evaluation will be performed.The reported event was unable to be confirmed or replicated.A cause of the reported event could not be determined at this time.A dhr review was unable to be completed as no lot information was provided.
 
Event or Problem Description
It was reported that th tracheostomy tube experienced a disconnection between the reinforced section and 30mm connector.This is an undetachable part of the tube, it appears to have become loose due to humidification.The tracheostomy tube had been in for 19 days.This resulted derecruitment of both lungs and desaturations, which was resolved by the placement of a new tracheostomy tube.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX UNIPERC TRACHEOSTOMY TUBES
Common Device Name
TUBE, TRACHEOSTOMY (W/WO CONNECTOR)
Manufacturer (Section D)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK  CT21 6JL
Manufacturer (Section G)
NULL
boundary road
hythe, kent CT21 6JL
UK   CT21 6JL
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key13542107
Report Number3012307300-2022-03460
Device Sequence Number3734497
Product Code BTO
UDI-Device Identifier15019315060441
UDI-Public15019315060441
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K083031
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number100/897/080
Device Catalogue Number100/897/080
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/10/2021
Supplement Date Received by Manufacturer06/14/2023
Initial Report FDA Received Date02/16/2022
Supplement Report FDA Received Date06/20/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-