The device was not returned for evaluation and the reported event could not be confirmed.The clinical/medical investigation concluded that, no further clinical information will be provided.Subsequently, without the requested relevant clinical information or the device, a thorough medical investigation cannot be rendered nor can a definitive root cause of the reported failure be determined.Therefore, the impact to the patient beyond that which has already been reported cannot be confirmed nor concluded.Should any additional relevant clinical information be provided, this case would be re-assessed.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history revealed similar events for the listed device over the previous 12 months, but no similar events for the batch based on the historical data, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A review of the instructions for use for this device reveal in the potential complications associated with total hip arthroplasty surgery, primary or revision section, that infection, both early, post-operative superficial and early, post-operative deep wound infection and late periprosthetic infection can occur.A review of the sterilization records revealed the batch was sterilized within normal parameters.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Possible causes could include but are not limited to contamination, patient reaction, and post-operative healing issue.The contribution of the device to the reported event could not be corroborated.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|