• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC IRELAND VALIANT NAVION; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC IRELAND VALIANT NAVION; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TR Back to Search Results
Model Number VNMC3737C90TU
Medical Device Problem Codes Leak/Splash (1354); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Anemia (1706); Unspecified Infection (1930); Sepsis (2067)
Date of Event 12/07/2020
Type of Reportable Event Death
Additional Manufacturer Narrative
Other relevant device(s) are: product id: vnmc3737c182tu, serial/lot #: (b)(4), ubd: 20-aug-2022, udi#: (b)(4).Product id: vnmc3737c90tu, serial/lot #: (b)(4), ubd: 20-aug-2022, udi#: (b)(4).If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Valiant navion stent grafts were implanted during the endovascular treatment of a thoracic aortic fusiform aneurysm.It was reported approximately 2 months post the index procedure follow-up ct identified a type ii endoleak.The patient was hemodynamically unstable and was admitted with a possible graft infection and anemia.An mra demonstrated an increased size of the aneurysm sac and a type iiia endoleak at the sma.The image showed a disconnection from the manifold.The infection led to sepsis and multi -organ failure and eventual death approximately 11 weeks post the index procedure. as per the physician the cause of the type ii and iiia endoleak are undetermined.The cause of death was due to hemodynamic instability with sepsis leading to multi-organ failure and death.No additional clinical sequelae was provided and the patient has expired.
 
Additional Manufacturer Narrative
It was confirmed the type iiia endoleak was between the manifold device and the non mdt sma bridging stents.The navion stent grafts had been implanted proximal to the manifold device.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
B5; additional information received: during the index procedure 2 endurant limbs (etlw1616c156e & etlw1624c156e) were implanted.An additional endurant limb (etlw1616c199e) was implanted bridging the valiant graft and visceral bypass of the ipsilateral limb of the bifurcated device was completed.2 non mdt stents were implanted in the leia and post dilated with a non mdt balloon.The max aortic diameter was noted as 107mm; the aneurysm size was noted to have had increased with the stent graft noted as patent.A type ii endoleak was observed and a type iiia endoleak was observed at the sma with a disconnection from the manifold.Non mdt stents were implanted from the manifold to the sma for the disconnection.A4.Updated b7.Updated medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VALIANT NAVION
Common Device Name
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TR
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
EI 
Manufacturer (Section G)
MEDTRONIC IRELAND
parkmore business park west
galway
EI  
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key13542321
Report Number9612164-2022-00640
Device Sequence Number14420060
Product Code MIH
UDI-Device Identifier00763000260842
UDI-Public00763000260842
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
P100040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/20/2022
Device Model NumberVNMC3737C90TU
Device Catalogue NumberVNMC3737C90TU
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/08/2022
Supplement Date Received by Manufacturer02/17/2022
03/16/2022
Initial Report FDA Received Date02/16/2022
Supplement Report FDA Received Date03/11/2022
04/07/2022
Date Device Manufactured08/20/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
"SEE H10...."
Outcome Attributed to Adverse Event Death;
Patient Age43 YR
Patient SexMale
Patient Weight143 KG
-
-