| Model Number |
826631 |
| Medical Device Problem Code |
Connection Problem (2900)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/27/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A physician reported that when the operator opened the microsensor ((b)(4)) and connected it to the main unit during the surgery, it was not recognized.The procedure was completed with a replacement product available.The issue was detected before the implantation site closure.No surgical delay or adverse consequences to the patient were observed.
|
| |
|
Additional Manufacturer Narrative
|
|
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
The microsensor (id: 826631) was not returned for evaluation after three good faith attempts (gfes) were made.Lot number information has not been provided; therefore, an evaluation of the device could not be performed, and device history record (dhr) could not be reviewed.The root cause(s) of the reported issue could not be determined.However, a probable root cause for the reported complaint is due to excessive pressure applied to product.If additional relevant information becomes available in the future, this complaint will be reopened, and the respective evaluation performed.Trends will be monitored for this and similar issues.At present, we consider this complaint to be closed.
|
| |
|
Event or Problem Description
|
|
N/a.
|
| |
|
Additional Manufacturer Narrative
|
|
The microsensor (id 826631) was returned for evaluation.Device history record (dhr) - for lot 3974832 (sn (b)(6)), and the lot met specifications when released.Failure analysis - the issue of the complaint was not confirmed.No visible damage to the millar sensor, catheter material, or connector.Cerelink monitor reading was acceptable; icp express passed with reading 491.The device passed electronic, noise, linearity/hysteresis, and signal drift tests.The root cause of the issue reported by customer could not be determined as the device worked correctly.However, the possible root cause of the defect reported by the customer could be due to excessive pressure applied to product.
|
| |
|