• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW MEDICAL LTD. ALLEVYN GENTLE BORDER 7.5X7.5CM INT; DRESSING,WOUND,OCCLUSIVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW MEDICAL LTD. ALLEVYN GENTLE BORDER 7.5X7.5CM INT; DRESSING,WOUND,OCCLUSIVE Back to Search Results
Model Number 66800269
Medical Device Problem Code Delivered as Unsterile Product (1421)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/28/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that an opening was observed on the packaging of one (1) allevyn gentle border 7.5cm x 7.5cm dressing.As this was noticed in a non-therapeutical environment, there was no patient involvement.
 
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event or Problem Description
It was reported that an opening was observed on the packaging of one (1) allevyn gentle border 7.5cm x 7.5cm dressing.As this was noticed in a non-therapeutical environment, there was no patient involvement.
 
Additional Manufacturer Narrative
H3, h6: the device was not used in treatment, has been returned and evaluated, establishing a relationship between the device and the reported event.Visual inspection observed the dressing was curled and cuptured at the edge of the seal.The root cause has been determined as a quality control deficiency.The dressing skewed after passing camera detection and then trapped in the side seal of pouch.The trapped dressing was not captured by operator during visual inspection.The batch record was reviewed and it could be confirmed that the products had progressed through a satisfactory release process which involved testing the product against the requirements of the finished product specification.There was also no issue identified during the manufacturing process that could have caused or contributed to the complaint problem.The complaint history review confirms previous complaints of this nature with manufacturing problems observed.A risk management review concluded that the alleged failure mode and any associated harm has been anticipated within the risk file, with no updates required.A review of prior escalation actions found no actions applicable to the scope of this case.This investigation is now complete with no further action deemed necessary.Smith and nephew will continue to monitor for any adverse trends relating to this product range.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALLEVYN GENTLE BORDER 7.5X7.5CM INT
Common Device Name
DRESSING,WOUND,OCCLUSIVE
Manufacturer (Section D)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull east riding of yorkshire HU3 2 BN
UK  HU3 2BN
Manufacturer (Section G)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull east riding of yorkshire HU3 2 BN
UK   HU3 2BN
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key13542956
Report Number8043484-2022-00071
Device Sequence Number3734518
Product Code NAD
UDI-Device Identifier05000223463878
UDI-Public05000223463878
Combination Product (Y/N)N
Initial Reporter CountryCO
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/01/2024
Device Model Number66800269
Device Catalogue Number66800269
Device Lot Number202112
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/28/2022
Supplement Date Received by Manufacturer07/19/2022
Initial Report FDA Received Date02/16/2022
Supplement Report FDA Received Date07/19/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/01/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-