Model Number 82-8814PL
Medical Device Problem Code Obstruction of Flow (2423)
Health Effect - Clinical Code Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 01/28/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
 
Event or Problem Description
A physician reported a certas valve (id 828814) was implanted via v-p shunt on (b)(6) 2022 with unknown setting.After the valve was implanted there was no change in the ventricles.On (b)(6) 2022 the set pressure could not be changed from setting 6 to 1.On (b)(6) 2022 another attempt was made to change the set pressure using a neodymium magnet, but the pressure could not be changed either.After adding a contrast medium and confirming patency, valve obstruction was confirmed.The valve was removed and replaced on (b)(6) 2022.
 
Additional Manufacturer Narrative
The certas valve (id 828814) was returned for evaluation.Failure analysis - the position of the cam when valve was received was at setting 6.The valve was visually inspected; a needle hole in the needle chamber were noted.The valve was hydrated.The valve was leak tested and only leaked from the needle hole in the needle chamber.The valve passed the test for programming, occlusion, reflux, siphon guard and pressure.No root cause could be determined as the technician could not confirm any problem with the valve at the time of investigation.The possible root cause for the issue reported by the customer could be due to biological debris and protein build up interfering with the valve mechanism, but at the time of investigation no occlusion issues were noted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CRTS SM IL SG VLV ONLY
Common Device Name
CERTAS PLUS SMALL & RA W/SG
Manufacturer (Section D)
INTEGRA LIFESCIENCES SWITZERLAND SAR
rue girardet 29
rue girardet 29
le locle
SZ 
Manufacturer (Section G)
INTEGRA LIFESCIENCES SWITZERLAND SAR
rue girardet 29
ch-2400
le locle
SZ  
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 
6099362319
MDR Report Key13543067
Report Number3013886523-2022-00053
Device Sequence Number4457385
Product Code JXG
UDI-Device Identifier10886704081937
UDI-Public10886704081937
Combination Product (Y/N)N
PMA/510(K) Number
NI
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number82-8814PL
Device Catalogue Number828814
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/28/2022
Supplement Date Received by Manufacturer03/28/2022
Initial Report FDA Received Date02/16/2022
Supplement Report FDA Received Date04/12/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexMale
-
-