MAUDE Adverse Event Report: VYAIRE MEDICAL INC. 3100 HIGH FREQUENCY OSCILLATING VENTILATOR (HFOV); VENTILATOR, HIGH FREQUENCY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VYAIRE MEDICAL INC. 3100 HIGH FREQUENCY OSCILLATING VENTILATOR (HFOV); VENTILATOR, HIGH FREQUENCY Back to Search Results
Model Number 3100A
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/19/2022
Type of Reportable Event Malfunction
Event or Problem Description
The customer reported the device will automatically start oscillating when the unit is turned on the 3100 a device.The customer reported there was no patient involvement associated with the reported event.
 
Additional Manufacturer Narrative
Vyaire medical file identification: (b)(4).At this time, the suspect device has not been returned for evaluation.Therefore, root cause has not been determined yet.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
 
Additional Manufacturer Narrative
Results of investigation: vyaire field service technician went onsite and evaluated the device.Technician installed the new power module.Issue was resolved.Performed 7 year + 12k preventive maintenance completed.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3100 HIGH FREQUENCY OSCILLATING VENTILATOR (HFOV)
Common Device Name
VENTILATOR, HIGH FREQUENCY
Manufacturer (Section D)
VYAIRE MEDICAL INC.
26125 n. riverwoods blvd.
mettawa IL 60045
Manufacturer (Section G)
VYAIRE MEDICAL INC.
1100 bird center drive
palm springs CA 92262
Manufacturer Contact
erika bonilla
510 technology dr
irvine, IL 92618
7149227837
MDR Report Key13544085
Report Number2021710-2022-15457
Device Sequence Number2187412
Product Code LSZ
UDI-Device Identifier10846446003024
UDI-Public(01)10846446003024(11)20080815
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
P890057
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number3100A
Device Catalogue Number768901-RFB
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/19/2022
Supplement Date Received by Manufacturer04/20/2022
Initial Report FDA Received Date02/16/2022
Supplement Report FDA Received Date05/09/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/15/2008
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-