Additional information was received 20apr2022.The groins were scarred bilaterally, resulting from a history of multiple endovascular and open vascular procedures.The patient was sedated, heparin was given, and local anesthesia was provided.Access was obtained in the left femoral artery, where a previous patch angioplasty had been performed, to target vessels of the right leg, below-the-knee.Advancement of the device into the groin and over the iliac bifurcation was reportedly difficult and resistance was reported, as the groin was scarred, and the iliac arteries were reportedly tortuous.Multiple wires, catheters, and balloons were passed through the sheath.Upon completion of the intended procedure, the sheath separated upon removal of the device over a wire, approximately twenty to thirty centimeters from the tip.The dilator was not reinserted prior to removal of the sheath and was not in place at the time of separation.The separated portion of the sheath (approximately twenty centimeters) remained in the left iliac artery.General anesthesia was administered, and the patient was intubated.The left femoral artery was exposed, and the sheath was retrieved.One unit of packed red blood cells was administered after the procedure.The patient was discharged from the hospital three days after the procedure with no medical complications.
|
|
Summary of event: as reported, during a procedure involving angioplasty with stent placement via a cross-over approach, a flexor high-flex ansel guiding sheath separated.The groins were scarred bilaterally, resulting from a history of multiple endovascular and open vascular procedures.The patient was sedated, heparin was given, and local anesthesia was provided.Access was obtained in the left femoral artery, where a previous patch angioplasty had been performed, to target vessels of the right leg, below-the-knee.Advancement of the device into the groin and over the iliac bifurcation was reportedly difficult and resistance was reported, as the groin was scarred, and the iliac arteries were reportedly tortuous and calcified.Multiple wires, catheters, and balloons were passed through the sheath.Upon completion of the intended procedure, the sheath separated upon removal of the device over a wire, approximately twenty to thirty centimeters from the tip.The dilator was not reinserted prior to removal of the sheath and was not in place at the time of separation.The separated portion of the sheath (approximately twenty centimeters) remained in the left iliac artery.General anesthesia was administered, and the patient was intubated.The left femoral artery was exposed, and the sheath was retrieved.One unit of packed red blood cells was administered after the procedure.The patient was discharged from the hospital three days after the procedure with no medical complications.Investigation evaluation: reviews of the complaint history, device history record, instructions for use (ifu), manufacturing instructions, and quality control procedures were conducted during the investigation.A visual inspection of the complaint device was also conducted.The used complaint device was returned to cook for investigation.Approximately eight centimeters of the material was separated from the sheath, with the inner coil exposed at the separation point.A document-based investigation evaluation was performed.No related non-conformances were found, and there have been no other reported complaints for this lot number.The product ifu warns that if resistance is encountered during sheath advancement, ¿assess cause of resistance and consider dilation of any restriction identified or consider alternate treatment strategy.If flexor sheath is advanced through resistance, force to remove the sheath will be higher, increasing the risk of sheath material or hub separation upon withdrawal¿.The ifu also states that reinsertion of the dilator prior to removal of the sheath ¿increases the strength of the sheath and lessens the risk of device separation¿ and suggests insertion of the dilator prior to removal if resistance is encountered or anticipated during withdrawal of the device.The ifu also cautions, ¿when puncturing, suturing or increasing the tissue near the sheath, use caution to avoid damaging the sheath.¿ the information provided upon review of the dmr, dhr, ifu, and investigation of the returned device suggests that there is evidence the device was manufactured to specification.There is no evidence of non-conforming devices in-house or in the field.Cook has concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.Based on the information provided and the results of the investigation, cook has concluded that operational factors contributed to this event, as the dilator was not reinserted into the sheath prior to removal of the device.It is also possible that the patient¿s anatomy contributed to the event, as the vessels were tortuous and calcified, the access site was scarred, and the patient had undergone multiple prior vascular procedures.The risk analysis for this failure mode was reviewed and no additional escalation was required.The appropriate personnel have been notified and cook will continue to monitor for similar events.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|