• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US SIGMA PS CEM FEM SZ4 R; SIGMA KNEE PRIMARY : KNEE FEMORAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US SIGMA PS CEM FEM SZ4 R; SIGMA KNEE PRIMARY : KNEE FEMORAL Back to Search Results
Model Number 1960-50-400
Medical Device Problem Code Loss of or Failure to Bond (1068)
Health Effect - Clinical Code Pain (1994)
Date of Event 02/04/2022
Type of Reportable Event Serious Injury
Event or Problem Description
Patient underwent revision due to painful knee.Femoral component was found to be loose at cement to implant interface during revision.Unknown cement used.All hardware were removed and revised to an attune revision femoral component with 16x60 stem and revision tibial component with s3 tibial sleeve and 16x60 stem and 12 mm poly.Doi: 2008; dor: (b)(6) 2022; affected side: right knee.
 
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot : a manufacturing record evaluation (mre), was not possible because the required lot code was not provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SIGMA PS CEM FEM SZ4 R
Common Device Name
SIGMA KNEE PRIMARY : KNEE FEMORAL
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kara ditty-bovard
700 orthopaedic dr.
warsaw, IN 46581-0988
6107428552
MDR Report Key13544220
Report Number1818910-2022-02955
Device Sequence Number5201415
Product Code JWH
UDI-Device Identifier10603295075141
UDI-Public10603295075141
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
P830055
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2008
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1960-50-400
Device Catalogue Number196050400
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/04/2022
Supplement Date Received by Manufacturer03/22/2022
Initial Report FDA Received Date02/16/2022
Supplement Report FDA Received Date03/24/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
BONE CEMENT (MFG. AND PRODUCT NAME UNKNOWN).; MBT CEM KEEL TIB TRAY SZ4.; SIG RP AOX STB INS SZ4 12.5.
Outcome Attributed to Adverse Event Required Intervention;
Patient Age83 YR
Patient SexMale
-
-