| Model Number |
TN DXH 800HEMATOLOGY SYSTEM WITH FLOOR STAND |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/24/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The customer technical support specialist (cts) recommended the customer perform the sweep flow prime procedure.The customer performed sweep flow prime three times resolving the event.Section a patient data was not provided.Bec internal identifier - case-(b)(4).
|
| |
|
Event or Problem Description
|
|
The customer reported patient samples recovered with high platelet (plt) results and no suspect messaging when cycled on their unicel dxh800 instrument compared to a rerun on an alternate instrument.Printouts of the event were requested, but not provided.There was no report of death, injury, or change to patient treatment as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
The customer technical support specialist (cts) recommended the customer perform the sweep flow prime procedure.The customer performed sweep flow prime three times resolving the event.Section a patient data was not provided.Bec internal identifier - (b)(4).
|
| |
|
Event or Problem Description
|
|
The customer reported patient samples recovered with high platelet (plt) results and no suspect messaging when cycled on their unicel dxh800 instrument compared to a rerun on an alternate instrument.Printouts of the event were requested, but not provided.There was no report of death, injury, or change to patient treatment as a result of this event.
|
| |
|