Model Number SB-0520FC
Medical Device Problem Code Peeled/Delaminated (1454)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/20/2022
Type of Reportable Event Malfunction
Event or Problem Description
As reported for this event by the customer, during a therapeutic tonsillectomy, after the device was activated two to three times, the tip of the ptfe pad (teflon pad) was partially peeled from the upper jaw.The teflon pad was burned overall as well.The procedure was completed with a new device.There is no reported harm or adverse impact to the patient.
 
Additional Manufacturer Narrative
Full name of the facility is the (b)(6) university of korea, (b)(6) hospital.The device is returned but the device evaluation is not yet completed.As such, a definitive root cause of the reported complaint cannot be determined at this time.Supplemental report(s) will be filed as any relevant new information becomes available.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation and device evaluation.The device was inspected and the event was confirmed; the tissue pad was severely worn out.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, it is likely the reported phenomenon occurred by the following mechanism: the ultrasonic output was activated while nothing was grasped between the grasping section and the distal end of the probe (this includes after tissue resection).Therefore, the tissue pad was worn out, and partially torn.This event can be prevented by following the instructions for use which state: "do not activate output while the grasping section is closed without contacting tissue or vessel, or ensuring that tissue is transected.Otherwise, a local increase of the temperature due to a friction between the probe tip and the grasping section may result in various forms of damage in the probe tip and/or the tissue pad, such as premature wear, breakage, deformation, and/or falling off inside the body cavity and/or partial separating." "when cutting or vessel sealing is performed, apply light tension on the tissue so that users can confirm that they are transected.Also, stop activation immediately after tissue is transected.Otherwise, the grasping section, the tissue pad, or the probe tip may break and fall, and partial separating of the tissue pad may occur due to a local increase of temperature caused by the friction between tissue pad and the probe tip during activation." olympus will continue to monitor field performance for this device.
 
Event or Problem Description
Addendum feb 17, 2022: the doctor is experienced in the use of the device.It is not known when during the procedure the event occurred.The settings were seal and cut mode 1.There was no delay in the completion of the procedure.
 
Additional Manufacturer Narrative
Additional information has been received from the customer.This supplemental report is being submitted to provide this information.The concomitant devices used in the procedure are not known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SONICBEAT 5 MM, 20 CM, FRONT-ACTUATED GRIP
Common Device Name
INSTRUMENT
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key13544420
Report Number8010047-2022-03064
Device Sequence Number16697516
Product Code LFL
UDI-Device Identifier04953170370601
UDI-Public04953170370601
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
K211838
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSB-0520FC
Device Lot Number0ZK02
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/06/2022
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/24/2022
Supplement Date Received by Manufacturer02/17/2022
02/22/2022
Initial Report FDA Received Date02/16/2022
Supplement Report FDA Received Date02/17/2022
03/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/02/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexPrefer Not To Disclose
-
-