Product was purchased on an order number prior to acquisition by symmetry surgical.A940 product is no longer available for sale.Customer stated they were completing a biopsy of skin on a patient with beard, when beard was caught on fire.Customer was contacted for information about the event, however, they were unwilling to answer any investigative questions.Complaint is unable to confirmed at this time.Root cause is unable to be determined with accuracy, however, it is believed this is a possible user error related event in not following the below precautions stated in a940 ifu, (b)(4).Fire / explosion hazard - the following substances will contribute to increased fire and explosion hazards in the operating room: flammable substances (hair is considered a flammable substance).The sparking and heating associated with electrosurgery can provide an ignition source.Observe fire precautions at all times.When using electrosurgery in the same room with any of these substances or gases, prevent their accumulation or pooling undersurgical drapes, or within the area where electrosurgery is performed.Based on the above information, this can be seen as the final report.If additional information is obtained that alleges any additional patient involvement or the need for corrective actions, a follow up report will be submitted.
|