• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DAVOL INC., SUB. C.R. BARD, INC. -1213643 VENTRALIGHT ST W/ ECHO 2; SURGICAL MESH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DAVOL INC., SUB. C.R. BARD, INC. -1213643 VENTRALIGHT ST W/ ECHO 2; SURGICAL MESH Back to Search Results
Catalog Number 5990015
Medical Device Problem Code Patient-Device Incompatibility (2682)
Health Effect - Clinical Code Bowel Perforation (2668)
Date of Event 01/27/2022
Type of Reportable Event Serious Injury
Event or Problem Description
As reported, during laparoscopic ventral hernia repair procedure on (b)(6) 2022, when the bard/davol ventralight st w/echo 2 mesh was inserted with the introducer tool, the introducer tool hit and nicked bowel as surgeon was pushing the introducer tool through the trocar at the incision site.As reported, the surgeon removed the introducer tool and did not place the bard/davol ventralight st w/echo 2 mesh, instead bard/davol phasix st echo 2 was used after the bowel was repaired using silk suture.
 
Additional Manufacturer Narrative
As reported, during insertion of the ventralight st w/echo ps 2 with introducer tool, the patient¿s bowel was injured when the bowel was inadvertently ¿nicked¿ with the introducer tool.Requests for additional information have been made to the user facility, however no additional information has been provided.Based on the event as reported, this event has a use related root cause as the surgeon inadvertently ¿nicked¿ the patient¿s bowel during insertion of the device.Review of manufacturing records confirms product was manufactured to specification, with no indication of a manufacturing related cause for the event reported.To date, this is the only reported complaint from this manufacturing lot of 115 units released for distribution in november, 2021.
 
Additional Manufacturer Narrative
As reported, during insertion of the ventralight st w/echo ps 2 with introducer tool, the patient¿s bowel was injured when the bowel was inadvertently ¿nicked¿ with the introducer tool.Requests for additional information have been made to the user facility, however no additional information has been provided.Based on the event as reported, this event has a use related root cause as the surgeon inadvertently ¿nicked¿ the patient¿s bowel during insertion of the device.Review of manufacturing records confirms product was manufactured to specification, with no indication of a manufacturing related cause for the event reported.To date, this is the only reported complaint from this manufacturing lot of (b)(4) units released for distribution in november, 2021.
 
Event or Problem Description
As reported, during laparoscopic ventral hernia repair procedure on (b)(6) 2022, when the bard/davol ventralight st w/echo 2 mesh was inserted with the introducer tool, the introducer tool hit and nicked bowel as surgeon was pushing the introducer tool through the trocar at the incision site.As reported, the surgeon removed the introducer tool and did not place the bard/davol ventralight st w/echo 2 mesh, instead bard/davol phasix st echo 2 was used after the bowel was repaired using silk suture.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VENTRALIGHT ST W/ ECHO 2
Common Device Name
SURGICAL MESH
Manufacturer (Section D)
DAVOL INC., SUB. C.R. BARD, INC. -1213643
100 crossings blvd.
warwick RI 02886
Manufacturer (Section G)
BARD SHANNON LIMITED -3005636544
san geronimo industrial park
lot #1, road #3, km 79.7
humacao PR 00791
Manufacturer Contact
andrew topoulos
100 crossings blvd.
warwick, RI 02886
8005566756
MDR Report Key13544622
Report Number1213643-2022-00019
Device Sequence Number19379864
Product Code FTL
UDI-Device Identifier00801741130991
UDI-Public(01)00801741130991
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K170294
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 01/31/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/28/2022
Device Catalogue Number5990015
Device Lot NumberHUFW0394
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/27/2022
Initial Report FDA Received Date02/16/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/02/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-