OLYMPUS MEDICAL SYSTEMS CORP. VISERA ELITE II XENON LIGHT SOURCE Back to Search Results
Model Number CLV-S200-IR
Medical Device Problem Code Electrical /Electronic Property Problem (1198)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/19/2022
Type of Reportable Event Malfunction
Event or Problem Description
A user facility reported to olympus that the light source was not working.Upon inspection of the device by a field service engineer, it was determined the emergency lamp indicator was blinking.The problem of "light source was not working," as reported to olympus, was identified during preparation for use.There was no patient injury, associated with the problem, reported to olympus.
 
Additional Manufacturer Narrative
The device was evaluated onsite at the user facility by a field service engineer (fse).The fse determined the emergency lamp indicator was blinking.The investigation is ongoing and the root cause of the reported event cannot be determined at this time.If additional information becomes available, this report will be supplemented accordingly.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information, based on the legal manufacturer's final investigation.A review of the device history record found no deviations, that could have caused or contributed to the reported issue.Based on the results of the legal manufacturer's investigation, the failure of the emergency light was confirmed, and was indicated by the flashing of the emergency light indicator on the front panel or by the display of error code e207.The root cause of the failure of the emergency lights could not be identified.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VISERA ELITE II XENON LIGHT SOURCE
Common Device Name
XENON LIGHT SOURCE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key13545037
Report Number8010047-2022-03074
Device Sequence Number17443445
Product Code OWN
UDI-Device Identifier04953170428142
UDI-Public04953170428142
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K200542
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 05/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCLV-S200-IR
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/19/2022
Supplement Date Received by Manufacturer04/28/2022
Initial Report FDA Received Date02/16/2022
Supplement Report FDA Received Date05/25/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/21/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-