• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. REDAPT STEM EXTRACTOR UNIV JOINT; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/POLYMER, CEMENTED OR NON-POROUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. REDAPT STEM EXTRACTOR UNIV JOINT; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/POLYMER, CEMENTED OR NON-POROUS Back to Search Results
Model Number 71355107
Medical Device Problem Codes Break (1069); Material Twisted/Bent (2981)
Health Effect - Clinical Codes Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/28/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the redapt stem extractor broke.It is unknown whether this problem was noticed in a surgical environment or not; therefore, patient involvement has not been confirmed.
 
Additional Manufacturer Narrative
Internal complaint reference case-(b)(4).
 
Additional Manufacturer Narrative
Internal complaint reference case-(b)(4).
 
Event or Problem Description
It was reported that the redapt stem extractor broke.It is unknown whether this problem was noticed in a surgical environment or not; therefore, patient involvement has not been confirmed.
 
Additional Manufacturer Narrative
The device was not returned for evaluation and the reported event could not be confirmed.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history revealed similar events for the listed device over the previous 12 months, but no similar events for the batch based on the historical data, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.The contribution of the device to the reported event could not be corroborated.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Event or Problem Description
It was reported that the redapt stem extractor broke.It is unknown whether this problem was noticed in a surgical environment or not; therefore, patient involvement has not been confirmed.The visual inspection of the returned device also noticed that the tip is bent.
 
Additional Manufacturer Narrative
H3, h6: the associated device was returned and evaluated.A visual inspection of the returned device confirms a large portion of the bushing broke off.The broken piece was not returned.Also the visual confirms the threaded tip of the device is bent.The device shows significant signs of wear and usage.A review of complaint history revealed similar events for the listed device over the previous 12 months, but no similar events for the batch based on the historical data, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.Assessment of historical escalated cases concluded that there are no prior actions related to this device and failure mode.As the device broke and it can no longer fit its purpose, the contribution of the device to the reported event could be corroborated.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REDAPT STEM EXTRACTOR UNIV JOINT
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/POLYMER, CEMENTED OR NON-POROUS
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key13545161
Report Number1020279-2022-00749
Device Sequence Number12021721
Product Code LZO
UDI-Device Identifier00885556223697
UDI-Public00885556223697
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K121627
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71355107
Device Catalogue Number71355107
Device Lot Number17APG0055
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/28/2022
Supplement Date Received by Manufacturer03/17/2022
11/15/2022
Initial Report FDA Received Date02/16/2022
Supplement Report FDA Received Date03/20/2022
11/17/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/12/2017
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-