Model Number H820G
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a patient underwent an unknown procedure on an unknown date and suture was used.During use on the patient it was noted that the needle broke off when removing from package.There were no patient consequences reported.Additional information was requested.
 
Additional Manufacturer Narrative
(b)(4).A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.The following information was requested, but was unavailable: the only information i received was that needle breaks off when removing from package.What was the procedure & event date? what was the initial procedure? what was used to complete the procedure? additional information was requested: please confirm if the device was used on the patient.Please confirm if the needle broke, or if the needle became detached from the suture.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
 
Additional Manufacturer Narrative
Product complaint # (b)(4) date sent to the fda: 3/14/2022 additional information: d9, h3, h6 additional information was requested, the following was obtained: the notes i have do not show this was used on patient.I do not know where it broke.H3 investigational summary: visual analysis of the returned sample determined that one opened sample of product code h820 was received for evaluation.The product code is single armed, and the needle was not received for evaluation.In addition, there is enough impression and insertion of the needle on the suture end.As part of our quality process, the manufacturing records of this lot-serial number were reviewed, and the manufacturing standards were met prior to the release of this batch.No conclusion could be reached as to what caused the reported complaint.It should be noted that as part of our quality process, the device is visually inspected and functionally tested during manufacturing to ensure the device meets the required specifications prior to shipment.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of post-market surveillance.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GUT PL UD 27IN 5-0 S/A FS-2
Common Device Name
SUTURE, ABSORBABLE, NATURAL
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.-JUAREZ
avenida de las torres 7125
col salvacar
cuidad juarez
MX  
Manufacturer Contact
elba bello
1000 route 202
raritan, NJ 08869
MDR Report Key13545815
Report Number2210968-2022-01131
Device Sequence Number13996805
Product Code GAL
UDI-Device Identifier10705031037762
UDI-Public10705031037762
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K946173
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative,Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2026
Device Model NumberH820G
Device Catalogue NumberH820G
Device Lot NumberRLMCPE
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/19/2022
Supplement Date Received by Manufacturer02/17/2022
Initial Report FDA Received Date02/16/2022
Supplement Report FDA Received Date03/14/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/07/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-