|
(b)(4).Date sent: (b)(6) 2022.D4 batch #: u95f06.Investigation summary: the product was returned to ethicon endo surgery for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that the hdh05 device was returned with the distal tip of the blade broken off and it was returned with the device.The remaining blade portion was scratched, and with evidence of contact with metal in or out of the operative field.During functional testing on gen11 an alert screen was displayed.A probable cause of the device stop activating and display an alert screen is blade damage.Probable causes of blade damage, including breakage, are external contact during pre-op or general use, blade contact with other devices, staples or clips during the procedure or using any means other than the blade wrench to attach or detach the blade.Once minor blade damage has occurred, subsequent activations may increase damage severity and result in alert screens, such as ¿tighten assembly¿, ¿blade error detected¿ or "relaxed pressure on blade" followed by a ¿replace instrument¿ screen later in the procedure.As part of ethicon endo surgery quality process all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.
|