| Model Number |
P12.5-80D305 |
| Medical Device Problem Codes |
Collapse (1099); Crack (1135)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Information was received that the patient had finished lengthening and was in the consolidation phase when the crown cracked and the nail collapsed.No patient adverse event was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Device has not been returned for evaluation; however device return is anticipated.No conclusions can be drawn at this time.If any additional information is provided, a supplemental report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
Visual inspection of the returned nail revealed, that the housing tube had cracks at the anti-rotation lug and the crown region.Based on visual inspection, the reported failure was confirmed.The x-ray image showed the anti-rotation lug (crown) was seated properly in the housing body.The iqc inspection data for the lots of anti-rotation lug and housing tube was reviewed and confirmed the parts met design specifications, per the engineering drawings.The work order was reviewed and confirmed, the device passed all inspections per the acceptance tests.The anti-rotation lug being cracked, indicates the type of breakage that can be a result of excess stress (torsional force) that is applied to the nail as a result of the patient¿s daily activities.
|
| |
|
Event or Problem Description
|
|
No additional information was provided.
|
| |
|