• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 POWER-TRIALYSIS SHORT-TERM CURVED EXTENSION DIALYSIS CATHETER 13F 15CM TL; CATHETER, HEMODIALYSIS, TRIPLE LUMEN, NON-IMPLANTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 POWER-TRIALYSIS SHORT-TERM CURVED EXTENSION DIALYSIS CATHETER 13F 15CM TL; CATHETER, HEMODIALYSIS, TRIPLE LUMEN, NON-IMPLANTED Back to Search Results
Model Number N/A
Medical Device Problem Code Failure to Infuse (2340)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/26/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The date of event was not provided by the complainant/reporter, the date reflected in this report is the date bd became aware of the event.The manufacturer has received the sample and will evaluate.Results are expected soon.A lot history review (lhr) review is not possible, as no manufacturing lot number has been provided by the complainant.
 
Event or Problem Description
It was reported a patient has clotted off two lines in the last 12 hours suspected due to the line curving and causing a clot.No other information was provided.This report addresses the first device.
 
Additional Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, applicable previous investigation(s), photo analysis and applicable fmea documents.Based on a review of this information, the following was concluded: the complaint that the tubing clotted off was inconclusive due to the sample condition.A photograph of the proximal end of a powertrialysis catheter was returned for evaluation.The photograph was framed so that only the extension legs and were visible.The extension legs were not clamped and red liquid, which appeared to be blood, was visible in the venous and arterial extension legs.It could not be determined from the photograph if the line was actively being used and/or if the fluid in the extension leg was clotted.The extension legs were pre-curved but the curves were straightened out in the photograph.No tight kinks or collapsing was seen in the extension leg tubing.It is possible that a restriction was present in the curved section of the catheter, but this could not be independently confirmed from the returned photograph.Other possible contributing factors for an occlusion could include patient physiology (e.G., stenosed vessel), catheter maintenance, positional occlusions, catheter collapse, or catheter kinks.H3 other text: evaluation findings are in section h11.
 
Event or Problem Description
It was reported a patient has clotted off two lines in the last 12 hours suspected due to the line curving and causing a clot.No other information was provided.This report addresses the first device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWER-TRIALYSIS SHORT-TERM CURVED EXTENSION DIALYSIS CATHETER 13F 15CM TL
Common Device Name
CATHETER, HEMODIALYSIS, TRIPLE LUMEN, NON-IMPLANTED
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
johanna de oliveira
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key13546043
Report Number3006260740-2022-00359
Device Sequence Number19386549
Product Code NIE
UDI-Device Identifier00801741075094
UDI-Public(01)00801741075094
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K133456
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number5615150
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/26/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/26/2022
Supplement Date Received by Manufacturer03/09/2022
Initial Report FDA Received Date02/16/2022
Supplement Report FDA Received Date03/18/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-