Model Number 802111
Medical Device Problem Code Output Problem (3005)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/26/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during use of the device for cardiopulmonary bypass (cpb), the level sensor detector malfunctioned.The surgical procedure was completed successfully.There was no delay, no blood loss, no adverse consequences to the patient.
 
Additional Manufacturer Narrative
This complaint is related to (b)(4) / mfr# (b)(4).
 
Event or Problem Description
Per clinical review: on (b)(6) 2021, the team experienced a problem with a level detector malfunction on their heart lung machine (hlm), while on cardiopulmonary bypass (cpb).This was the second occasion of a similar malfunction.The manufacturer's clinical specialist spoke with the perfusionist and he stated that the level was giving a false notification, as the actual level in the venous reservoir was well above the sensor level.He eventually turned off the level detection system and completed the case successfully, with no delay, blood loss and no adverse event.
 
Additional Manufacturer Narrative
The field service representative (fsr) did not find any issues with the red level sensor.He did replace the level sensor module and the yellow level sensor.The level sensor module was part number: 802111, serial number: (b)(6) , udi: (b)(4).The yellow level sensor was part number: 195215, serial number: (b)(6) , udi: (b)(4).The unit operated to the manufacturer's specifications.
 
Additional Manufacturer Narrative
The reported complaint was confirmed.It was determined to update this complaint to be written on the level module since that unit was returned for evaluation rather than the level sensor which was determined by the field service representative (fsr) to operate to the manufacturer's specification.During laboratory analysis, the product surveillance technician (pst) connected the level detector module to lab use only (luo) testing equipment.The unit ran for 3.68 hours testing the alert/alarm system with no observation of any level detector malfunction.It was determined that the level detector module met specification.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANCED PERFUSION SYSTEM 1
Common Device Name
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
Manufacturer (Section G)
SAME
Manufacturer Contact
douglas patton
6200 jackson road
ann arbor, MI 48103
7346634145
MDR Report Key13546140
Report Number1828100-2022-00061
Device Sequence Number5201476
Product Code DTQ
UDI-Device Identifier00886799000717
UDI-Public(01)00886799000717(11)210311
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
K172220
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number802111
Device Catalogue Number802111
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/27/2022
Supplement Date Received by Manufacturer02/25/2022
03/28/2022
Initial Report FDA Received Date02/16/2022
Supplement Report FDA Received Date03/21/2022
04/14/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/11/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-