Model Number 41173E
Medical Device Problem Codes Leak/Splash (1354); Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/07/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Fda notified?: the initial reporter also notified the fda via medwatch # mw5106599.A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event or Problem Description
It was reported that the bd alaris smartsite gravity set separated and leaked.The following information was provided by the client: "the iv tubing just fell off of the drip chamber.".
 
Event or Problem Description
It was reported that the bd alaris smartsite gravity set separated and leaked.The following information was provided by the client: "the iv tubing just fell off of the drip chamber.".
 
Additional Manufacturer Narrative
H.6.Investigation: no photo or physical sample was received for the investigation into the separation of the 41173e infusion set.Without further evidence, the complaint cannot be verified and the root cause remains unknown.A device history record review for model 41173e lot number 21079170 was performed.There were no quality notifications issued for the failure mode reported by the customer during the build of this set.A complaint history check was performed and this is the 1st related complaint reported with the defect/condition of separation- leak regarding item #41173e over the 12 month period before this complaint.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD ALARIS SMARTSITE GRAVITY SET
Common Device Name
INTRAVASCULAR ADMINISTRATION SET
Manufacturer (Section D)
SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer (Section G)
SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key13546443
Report Number9616066-2022-00137
Device Sequence Number5125626
Product Code FPA
UDI-Device Identifier37613203020900
UDI-Public37613203020900
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K820278
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/26/2024
Device Model Number41173E
Device Catalogue Number41173E
Device Lot Number21079170
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Event Location Hospital
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/03/2022
Supplement Date Received by Manufacturer06/03/2022
Initial Report FDA Received Date02/16/2022
Supplement Report FDA Received Date06/08/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-