Model Number MD-631
Medical Device Problem Code Electrical /Electronic Property Problem (1198)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/23/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device was received and evaluated at olympus service center (b)(4).Device inspection, the following were noted during device evaluation: return equipment: lot is 15r.Visual inspection: there was no abnormality in the appearance.Actual machine confirmation: the problem was not reproduced.Operation check: there is no abnormality when connecting and operating according to the instruction manual.Based on evaluation findings, the reported issue was not able to be reproduced.Investigation is ongoing.This report will be supplemented accordingly following completion of investigation.
 
Event or Problem Description
It was reported that the user is using otv-s190 and clv-s190 in combination.A clv lamp lighting failure error e103 occurred when power is turned on.The user power off and turn on the device and when it starts up without any problem, an error e102 occurred.The user replaced the lamp with a spare lamp and the error disappeared.The customer stated that the subject lamp has been used for about 100 hours.The issue found at preparation for use.There is no patient harm or injury reported due to the event.No user injury reported.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.H4 - the device was manufactured in june 2016.A final device manufacturer date is not available.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the specific root cause of the errors codes could not be determined at this time as the event was not replicated during evaluation.The following information is stated in the instructions for use: "code: e102 a84.Error message: clv lamp goes out check the check lamp.Cause: the light source device is out of order.Corrective action: immediately turn off the power of the light source device and inspect the lighting lamp according to the "instruction manual" of the light source device.If there is nothing wrong with the lighting lamp, immediately turn off the power of the light source device, unplug the power plug of the light source device from the medical outlet, disconnect the power cord from the power inlet of the light source device, and contact olympus.Code: e103 a83.Error message: clv lamp lighting failure check the check lamp.Cause: the light source device is out of order.Corrective action: immediately turn off the power of the light source device and inspect the lighting lamp according to the "instruction manual" of the light source device.If there is nothing wrong with the lighting lamp, immediately turn off the power of the light source device, unplug the power plug of the light source device from the medical outlet, disconnect the power cord from the power inlet of the light source device, and contact olympus.".Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XENON LAMP
Common Device Name
XENON LAMP
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key13546550
Report Number8010047-2022-03103
Device Sequence Number20147809
Product Code GCT
UDI-Device Identifier04953170355950
UDI-Public04953170355950
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
CLASS2-EXMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date (Section B) 03/31/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMD-631
Device Lot Number15R
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/26/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/23/2022
Supplement Date Received by Manufacturer03/03/2022
Initial Report FDA Received Date02/16/2022
Supplement Report FDA Received Date03/31/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
OTV-S190. CLV-S190.
Patient SexUnknown
-
-