We checked the returned unit and confirmed that the ccd wire was broken.Based on the result, we concluded that it was caused due to the excessive force applied on the ccd wire.In addition, we confirmed that the u/d knob broken, the light guide cable fluid damage, the lg cable connector fluid damage, the lg cable connector corroded, the light guide cable for control body fluid damage, the shield cover corroded, and the lg cable connector housing scratched; however, they are not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report and/or the risk analysis results, it was evaluated to submit mdr.
|