BARD PERIPHERAL VASCULAR, INC. CROSSER; RECANALIZATION CATHETER Back to Search Results
Model Number CRU14SA
Medical Device Problem Codes Melted (1385); Material Puncture/Hole (1504); Device-Device Incompatibility (2919)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/19/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a recanalization procedure in a peripheral arterial disease of superficial femoral artery, the catheter was allegedly jammed on to the guidewire.It was further reported that the catheter was allegedly melted.The procedure was completed using wire escalation technique.There was no reported patient injury.
 
Additional Manufacturer Narrative
As the lot number for the device was provided, a review of the device history record is currently being performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.(expiry date: 10/2023).
 
Additional Manufacturer Narrative
As the lot number for the device was provided, a review of the device history record is currently being performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.(expiry date: 10/2023).
 
Event or Problem Description
It was reported that during a recanalization procedure in a peripheral arterial disease of superficial femoral artery, the catheter was allegedly jammed on to the guidewire.It was further reported that the catheter was allegedly melted.The procedure was completed using wire escalation technique.There was no reported patient injury.
 
Additional Manufacturer Narrative
H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: one 14s crosser cto recanalization catheter was returned for evaluation.Sample was received loaded onto an unknown guidewire.No burnt residue noted to the sample.The guidewire was removed from the catheter with no issues.Catheter puncture noted.Based on the findings, the investigation was confirmed for the identified material puncture issue.However, the investigation was unconfirmed for the reported device-device incompatibility and melted issues.A definitive root cause for the reported device-device incompatibility, melted issues and identified material puncture issue could not be determined based upon the provided information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4 (expiry date: 10/2023), g3, h6 (device, method).H11: h6 (result, conclusion).H11: section a through f ¿ the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : see h10.
 
Event or Problem Description
It was reported that during a recanalization procedure for a peripheral arterial disease of superficial femoral artery, the catheter was allegedly jammed on to the guidewire.It was further reported that the catheter was allegedly melted.The procedure was completed using wire escalation technique.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CROSSER
Common Device Name
RECANALIZATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
Manufacturer (Section G)
C.R. BARD, INC. (GFO)
289 bay road
queensbury 12804
Manufacturer Contact
judy ludwig
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key13547425
Report Number2020394-2022-00108
Device Sequence Number12058801
Product Code PDU
UDI-Device Identifier10801741125567
UDI-Public(01)10801741125567
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
K161208
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCRU14SA
Device Catalogue NumberCRU14SA
Device Lot NumberGFFX1249
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/10/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/21/2022
Supplement Date Received by Manufacturer03/22/2022
Initial Report FDA Received Date02/17/2022
Supplement Report FDA Received Date03/24/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/23/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1