| Model Number |
N/A |
| Medical Device Problem Code |
Component Missing (2306)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/31/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The user facility reported that there was no polytetrafluoroethylene (ptfe) coating on the guidewire involved.The patient was ok.The outcome of the procedure was positive.There was no patient injury/medical or surgical intervention required.Additional information was received on 01 feb 2022: the procedure was completed using a cook roadrunner wire.
|
| |
|
Additional Manufacturer Narrative
|
|
Implanted date: device was not implanted.Explanted date: device was not explanted.Occupation: leah tech.A review of the device history record of the product code/lot# combination was conducted with no findings.The actual device has not been returned for evaluation.The investigation is currently ongoing.A follow-up report will be submitted once the investigation is complete.(b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted as follow up no.2 to update section h3, and to provide the completed investigation results.Based upon the investigation of the unopened samples this event has been reassessed and has now been deemed not a reportable event.The inspection of the two unopened returned samples upon receipt found no abnormalities in the appearance, dimensions, or manufacturing related records of the actual sample.The cause of the problem reported could not be clarified snice the samples were verified to be of normal products; therefore, is not a reportable event.Ashitaka factory is always continuing diligence in maintaining the product quality by performing thorough work and inspections, for example, as follows: (i) the outer diameter of the product is controlled; (ii) the homogeneity of the hydrophilic coating solution is guaranteed by controlling the balance of materials and the stirring time; (iii) the uniformity of the hydrophilic coating layer is guaranteed by controlling the application conditions (coating speed and immersion time); (iv) after the application of hydrophilic coating, sliding performance is checked to ensure that the hydrophilic coating has been applied properly.Terumo medical products (tmp) (importer) registration no.2243441 is submitting this report on behalf of ashitaka factory of terumo corporation (manufacturer) registration no.9681834.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted as follow up no.1 to provide the device return date in section d9 and to update section h3.The investigation is currently ongoing.A follow-up report will be submitted once the investigation is complete.
|
| |
|