• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US SOL SYS 10/13.5 R BOW; SOLUTION STEM IMPLANTS : HIP FEMORAL STEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US SOL SYS 10/13.5 R BOW; SOLUTION STEM IMPLANTS : HIP FEMORAL STEM Back to Search Results
Model Number 1572-11-136
Medical Device Problem Code Migration (4003)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 01/31/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient had a periprosthetic of their competitor hemi hip prosthesis.The surgeon removed the indwelling hip and replaced it with a 10¿ bowed solution stem and a new hemi head/taper spacer on (b)(6) 2022.A few days after this surgery, the patient's hip dislocated.The surgeon determined that the stem rotated and needed to be replace with a larger diameter stem.This new procedure happened on (b)(6) 2022.The stem, head and taper spacer were all replaced.It was also indicated that there was loosening of the stem at the bone to implant interface.Doi: (b)(6) 2022, dor: (b)(6) 2022, affected side: right hip.
 
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that the patient had a periprosthetic of their competitor hemi hip prosthesis.The surgeon removed the indwelling hip and replaced it with a 10¿ bowed solution stem and a new hemi head/taper spacer on (b)(6) 2022.A few days after this surgery, the patient's hip dislocated.The surgeon determined that the stem rotated and needed to be replace with a larger diameter stem.This new procedure happened on (b)(6) 2022.The stem, head and taper spacer were all replaced.It was also indicated that there was loosening of the stem at the bone to implant interface.Doi: (b)(6) 2022, dor: (b)(6) 2022, affected side: right hip.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SOL SYS 10/13.5 R BOW
Common Device Name
SOLUTION STEM IMPLANTS : HIP FEMORAL STEM
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kara ditty-bovard
700 orthopaedic dr.
warsaw, IN 46581-0988
6107428552
MDR Report Key13548474
Report Number1818910-2022-02964
Device Sequence Number5199857
Product Code LPH
UDI-Device Identifier10603295061717
UDI-Public10603295061717
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K060581
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1572-11-136
Device Catalogue Number157211136
Device Lot NumberJK0304
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/31/2022
Supplement Date Received by Manufacturer03/15/2022
Initial Report FDA Received Date02/17/2022
Supplement Report FDA Received Date03/17/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/09/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
MODULAR CATHCART BALL 44MM OD; SOL SYS 10/13.5 R BOW; TAPERED SPACER ARTICUL/EZE +0
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-