DEPUY ORTHOPAEDICS INC US SIGMA HP UNI INS SZ2 8MM LM/RL; EARLY INTERVENTION : KNEE TIBIAL INSERT Back to Search Results
Model Number 1024-53-208
Medical Device Problem Code Difficult to Insert (1316)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 02/02/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that the sigma partial knee insert doesn't fit with tibia tray.Doctor cannot inserted to tibia.There was a ten minute surgical delay.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot : a manufacturing record evaluation (nc search) was performed for the finished device 102453208 , lot j8783m, and no non-conformances / manufacturing irregularities were identified.
 
Event or Problem Description
A.Please provide the hospital or account name.Answer: (b)(6) hospital b.It was indicated in the event description that ''early loosening / unstable.'' please clarify if these were the adverse consequences the patient experienced due to knee insert doesn't fit with tibia tray.Answer: normally when surgeon put insert to tibia it should be easy to put it to tibia but this case it¿s hard to put it and surgeon tried to put many times.He observed insert has unusual appearance.He think it¿s cause insert doesn't fit with tibia tray.B.Did the event happen during primary surgery or revision surgery? answer: primary c.If the event happened during a revision surgery, were there depuy products explanted? if yes, was the revision being reported? please provide the reference number.If none reported, please state the reason for revision surgery, product and lot number of all explanted depuy products.Additional information added: 1.Please confirm what the resolution was to the implant to implant fit issues.Answer: change size.
 
Additional Manufacturer Narrative
Product complaint #
=
> (b)(4) this report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: added: b5 and d10 if information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SIGMA HP UNI INS SZ2 8MM LM/RL
Common Device Name
EARLY INTERVENTION : KNEE TIBIAL INSERT
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kara ditty-bovard
700 orthopaedic dr.
warsaw, IN 46581-0988
6107428552
MDR Report Key13548522
Report Number1818910-2022-02967
Device Sequence Number14733818
Product Code JWH
UDI-Device Identifier10603295002499
UDI-Public10603295002499
Combination Product (Y/N)N
Initial Reporter CountryTH
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2025
Device Model Number1024-53-208
Device Catalogue Number102453208
Device Lot NumberJ8783M
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/02/2022
Supplement Date Received by Manufacturer02/26/2022
03/18/2022
Initial Report FDA Received Date02/17/2022
Supplement Report FDA Received Date03/03/2022
03/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/12/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SIGMA HP UNI INS SZ2 8MM LM/RL; SIGMA HP UNI INS SZ2 8MM LM/RL; SIGMA HP UNI INS SZ2 8MM LM/RL; UNKNOWN KNEE TIBIAL TRAY
Patient SexUnknown
-
-